ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-02382
- Event Type
- Malfunction
- Date Received
- March 5, 2013
- Report Date
- February 7, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
FOLLOW-UP # 1 (B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY REVEALED MULTIPLE MANUAL TIME AND DATE CHANGES AROUND THE TIME OF THE EVENT. A TIMEKEEPING ACCURACY TEST WAS PERFORMED AND THE PUMP MAINTAINED THE TIME AND DATE ACCURATELY. THE COMPLAINT COULD NOT BE DUPLICATED.
ON (B)(4) 2013, THE REPORTER CONTACTED ANIMAS AND STATED HIS DAUGHTER HAD BEEN HAVING ERRATIC BLOOD GLUCOSE (BG) LEVELS FOR 1-2 WEEKS. AT TIME OF CALL BG IS IN 400 MG/DL RANGE. DAD HAS CONTACTED THE HEALTH CARE PROVIDER (HCP) AND EDUCATOR REGARDING THE SITES. DAD JUST WANTS TO TROUBLESHOOT THE PUMP TO BE SURE THE PUMP IS DELIVERING AS INTENDED. DAUGHTER'S BGS VARIED AS FOLLOWS: (B)(6) 2013 AT 1:30 / 457, 10:30 AM / 372, 8:26 AM / 255, 6:27 AM / 268, 2:30 AM / 479; (B)(6) 2013 9:14 PM / 112, 7:30 PM / 145, 4:45 PM / 386, 1:01 PM / 267, 10 AM / 267 AND 7:05 AM / 169. DAD ADAMANT DAUGHTER IS CHANGING SITE AS INSTRUCTED. SHE IS ROTATING SITES ON ABDOMEN, BUT DAD FEELS SHE MAY BE STAYING IN THE SAME SMALL AREA. DAD IN A HURRY, AGAIN STATES JUST WANTS TO GO THROUGH THE PUMP. HAD DAD REMOVE THE BATTERY AND RESET PUMP, TIME/DATE MAINTAINED CORRECT. CONFIRMED ALL SETTINGS CORRECT IN PUMP, PRIME HISTORY CORRECT, PRIME CORRECT, BUT TIME DATE WAS OFF ON (B)(6). TDD HISTORY INDICATES 2 DATES OF(B)(6), WHICH THE PATIENT IS UNABLE TO EXPLAIN. THEN ON (B)(6) AM/PM TIME WAS TRANSPOSED, READ 4:50P -0-, 5P/1.1, 11A/1.O, AND THEN BACK TO 5P/1.1, BOLUS SKIPPED FROM (B)(6) AND BACK TO (B)(6) AGAIN. BOLUS CORRECT. ALL OTHER SETTINGS WERE CORRECT. THE PATIENT IS ADAMANT SHE DID NOT TOUCH THE TIME /DATE. THE BLOOD GLUCOSE EXCURSIONS DO NOT MEET THE CRITERIA FOR AN ADVERSE EVENT, AND THIS COMPLAINT IS BEING REPORTED DUE TO THE UNEXPLAINED DISCREPANCY IN DATE AND TIME ON (B)(6), WHICH MAY HAVE ALTERED INSULIN DELIVERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 93079 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |