FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2990280 · Received March 5, 2013

Report

Report Number
2531779-2013-02382
Event Type
Malfunction
Date Received
March 5, 2013
Report Date
February 7, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY REVEALED MULTIPLE MANUAL TIME AND DATE CHANGES AROUND THE TIME OF THE EVENT. A TIMEKEEPING ACCURACY TEST WAS PERFORMED AND THE PUMP MAINTAINED THE TIME AND DATE ACCURATELY. THE COMPLAINT COULD NOT BE DUPLICATED.

Description of Event or Problem · 1

ON (B)(4) 2013, THE REPORTER CONTACTED ANIMAS AND STATED HIS DAUGHTER HAD BEEN HAVING ERRATIC BLOOD GLUCOSE (BG) LEVELS FOR 1-2 WEEKS. AT TIME OF CALL BG IS IN 400 MG/DL RANGE. DAD HAS CONTACTED THE HEALTH CARE PROVIDER (HCP) AND EDUCATOR REGARDING THE SITES. DAD JUST WANTS TO TROUBLESHOOT THE PUMP TO BE SURE THE PUMP IS DELIVERING AS INTENDED. DAUGHTER'S BGS VARIED AS FOLLOWS: (B)(6) 2013 AT 1:30 / 457, 10:30 AM / 372, 8:26 AM / 255, 6:27 AM / 268, 2:30 AM / 479; (B)(6) 2013 9:14 PM / 112, 7:30 PM / 145, 4:45 PM / 386, 1:01 PM / 267, 10 AM / 267 AND 7:05 AM / 169. DAD ADAMANT DAUGHTER IS CHANGING SITE AS INSTRUCTED. SHE IS ROTATING SITES ON ABDOMEN, BUT DAD FEELS SHE MAY BE STAYING IN THE SAME SMALL AREA. DAD IN A HURRY, AGAIN STATES JUST WANTS TO GO THROUGH THE PUMP. HAD DAD REMOVE THE BATTERY AND RESET PUMP, TIME/DATE MAINTAINED CORRECT. CONFIRMED ALL SETTINGS CORRECT IN PUMP, PRIME HISTORY CORRECT, PRIME CORRECT, BUT TIME DATE WAS OFF ON (B)(6). TDD HISTORY INDICATES 2 DATES OF(B)(6), WHICH THE PATIENT IS UNABLE TO EXPLAIN. THEN ON (B)(6) AM/PM TIME WAS TRANSPOSED, READ 4:50P -0-, 5P/1.1, 11A/1.O, AND THEN BACK TO 5P/1.1, BOLUS SKIPPED FROM (B)(6) AND BACK TO (B)(6) AGAIN. BOLUS CORRECT. ALL OTHER SETTINGS WERE CORRECT. THE PATIENT IS ADAMANT SHE DID NOT TOUCH THE TIME /DATE. THE BLOOD GLUCOSE EXCURSIONS DO NOT MEET THE CRITERIA FOR AN ADVERSE EVENT, AND THIS COMPLAINT IS BEING REPORTED DUE TO THE UNEXPLAINED DISCREPANCY IN DATE AND TIME ON (B)(6), WHICH MAY HAVE ALTERED INSULIN DELIVERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93079 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 15 YR