LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2013-01046
- Event Type
- Injury
- Date Received
- March 5, 2013
- Date of Event
- February 18, 2013
- Report Date
- February 25, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). RESULTS: THE DEVICE WAS RETURNED FOR ANALYSIS AND UPON INSPECTION THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. PHOTOGRAPHIC ANALYSIS: THE TEAR DROP/PEAR SHAPE CLIP FORMATION WAS THE RESULT OF EXCESSIVE DOWNWARD FORCE PUSHING THE CLIP BACK INTO THE JAWS DURING FIRING. THE ISSUE AROUND ONE CLIP LEG LIFTING OUT OF THE CLIP TRACK AND NOT THE OTHER DURING FIRING IS NORMAL AND WHAT I WOULD EXPECT TO SEE IF ONE CLIP LEG IS RESTRICTED SLIGHTLY AT THE VERY BEGINNING OF THE FIRING STROKE.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, SURGEON COMPLAINED THAT THE CLIPS FROM THE DEVICE WERE NOT APPROXIMATING WHEN FIRED UPON THE CYSTIC DUCT. THE SURGEON ALSO TOOK PICTURES OF THE PROXIMAL END OF THE CLIP LIFTING FROM ONE SIDE OF THE JAWS WHEN BEGINNING THE FIRING PROCESS. THE CASE WAS COMPLETED USING AN EL5ML CLIP APPLIER AND THERE WERE NO PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 92963 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | K4C293 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |