FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2990222 · Received March 5, 2013

Report

Report Number
1416980-2013-05289
Event Type
Malfunction
Date Received
March 5, 2013
Date of Event
February 1, 2013
Report Date
February 14, 2013
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION:THE REPORTED CONDITION OF FAILURE CODE 38 WAS CONFIRMED DURING DEVICE EVALUATION. THE ASSIGNABLE CAUSE WAS IDENTIFIED AS WITH THE FORCE SENSING RESISTORS (FSRS). THE FSRS WERE REPLACED TO CORRECT THE REPORTED CONDITION. ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON-SITE AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FLOGARD PUMP REPORTED THAT THE EQUIPMENT PRESENTED THE ALARM F-38. NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION WERE REPORTED WHEN THIS EVENT WAS RECEIVED. ADDITIONAL INFORMATION IS NOT EXPECTED AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93762 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1