FDA Adverse Event Injury Summary report: N

ARTICULEZE M HEAD 36MM -2

MDR report key: 2990129 · Received March 5, 2013

Report

Report Number
1818910-2013-13207
Event Type
Injury
Date Received
March 5, 2013
Date of Event
February 21, 2013
Report Date
July 25, 2013
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
JDI
PMA / PMN Number
PK980513
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPDATE: (B)(6) 2013 - LITIGATION PAPERS RECEIVED. LITIGATION ALLEGES DISCOMFORT. THERE IS NO ADDITIONAL INFORMATION THAT WOULD AFFECT THE OUTCOME OF THIS INVESTIGATION.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

**UPDATE** (B)(4) 2013- PFS AND MEDICAL RECORDS RECEIVED. THE INVESTIGATION IS ONGOING. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

UPDATED: PATIENT WAS REVISED TO ADDRESS PAIN AND ELEVATED ION LEVELS. DOI (B)(6) 2010 - DOR (B)(6) 2013 (RIGHT HIP). THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE FEMORAL HEAD'S PRODUCT AND LOT COMBINATION SINCE ITS RELEASE FOR DISTRIBUTION. A PREVIOUS REVIEW OF THE DEVICE HISTORY RECORDS FOR THE ACETABULAR LINER'S PROVIDED PRODUCT/LOT COMBINATION DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. D. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE FOUND NO ADDITIONAL RELATED REPORTS FOR THE LOT CODE 3041445. A SEARCH OF THE COMPLAINT DATABASE FINDS ADDITIONAL REPORTED INCIDENTS AGAINST LOT CODE 3025399 SINCE ITS RELEASE FOR DISTRIBUTION; HOWEVER, A PREVIOUS REVIEW OF THE DEVICE HISTORY RECORDS DID NOT REVEAL ANY RELATED MANUFACTURING ANOMALIES. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS PAIN AND ELEVATED ION LEVELS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
94356 ARTICULEZE M HEAD 36MM -2 FEMORAL HEAD JDI DEPUY ORTHOPAEDICS INC US 3041445

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention