FDA Adverse Event Malfunction Summary report: N

HANSSON TWIN HOOK 90MM, OMEGA PLUS TI

MDR report key: 2990072 · Received March 5, 2013

Report

Report Number
0008031020-2013-00038
Event Type
Malfunction
Date Received
March 5, 2013
Date of Event
February 8, 2013
Report Date
February 8, 2013
Manufacturer
STRYKER OSTEOSYNTHESIS-SELZACH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED INCIDENT COULD NOT BE CONFIRMED, SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. HOWEVER, THE SURGEON REPORTED THAT HE THOUGHT THE PATIENTS' BONE QUALITY INFLUENCED. THEREFORE, A PATIENT DEVICE INTERFACE ISSUE COULD NOT BE EXCLUDED. INDICATIONS FOR ANY MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE NOT DETERMINED IN THE INVESTIGATION. PLEASE NOTE THAT MORE DETAILED INFORMATION ABOUT THE COMPLAINT EVENT AS WELL AS THE AFFECTED DEVICE MUST BE AVAILABLE IN ORDER TO DETERMINE THE ROOT CAUSE OF THE COMPLAINT EVENT. IF ANY FURTHER INFORMATION IS PROVIDED, THE INVESTIGATION REPORT WILL BE UPDATED. DEVICE NOT RETURNED.

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

DURING OMEGA TWIN HOOK SURGERY, THE SURGEON PUSHED THE HOOK OUT BY THE T-HANDLE. AFTERWARDS, WHEN THE SURGEON CONFIRMED BY X-RAY, THE HOOK WAS PUSHED OUT WITHOUT CURVING. THEREFORE, THE SURGEON PULLED BACK THE HOOK AND INSERTED THE NEW PIN. HOWEVER, THE NEW PIN WAS ALSO PUSHED OUT, WITHOUT CURVING. THE SURGERY WAS COMPLETED. THE SURGEON THINKS THAT THE PATIENT'S BONE QUALITY INFLUENCED THE SURGERY.

Description of Event or Problem · 1

DURING OMEGA TWIN HOOK SURGERY, THE SURGEON PUSHED THE HOOK OUT BY THE T-HANDLE. AFTERWARDS, WHEN THE SURGEON CONFIRMED BY X-RAY, THE HOOK WAS PUSHED OUT WITHOUT CURVING. THEREFORE, THE SURGEON PULLED BACK THE HOOK AND INSERTED THE NEW PIN. HOWEVER, THE NEW PIN WAS ALSO PUSHED OUT, WITHOUT CURVING. THE SURGERY WAS COMPLETED. THE SURGEON THINKS THAT THE PATIENTS' BONE QUALITY INFLUENCED THE SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93133 HANSSON TWIN HOOK 90MM, OMEGA PLUS TI INSTRUMENT LXH STRYKER OSTEOSYNTHESIS-SELZACH

Patients

Seq Age Sex Outcome Treatment
1 40 YR Other