ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-02375
- Event Type
- Malfunction
- Date Received
- March 5, 2013
- Date of Event
- February 13, 2013
- Report Date
- February 13, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWED MULTIPLE OCCLUSIONS. DURING THE LOAD STEP, THE PUMP RECOGNIZED THE CARTRIDGE WHEN NO CARTRIDGE WAS PRESENT AND THE PUMP EMITTED AN OCCLUSION ALARM WITHIN 3 MINUTES OF PRIMING. THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION AND THE FORCE RESISTANCE MEASUREMENT WAS OUT OF SPECIFICATION. THE PUMP WAS OPENED AND A FORCE SENSOR CIRCUIT COMPONENT WAS FOUND TO BE MISALIGNED ON THE PRINTED CIRCUIT BOARD. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP WAS RETURNED FOR LARGE PRIME VOLUME. INVESTIGATION REVEALED THAT A FORCE SENSOR CIRCUIT COMPONENT WAS MISALIGNED ON THE PRINTED CIRCUIT BOARD. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 93059 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR |