FDA Adverse Event Other Summary report: N

RICHARD ALLEN SURGICAL NEEDLE

MDR report key: 2990035 · Received February 21, 2013

Report

Report Number
1836161-2013-00022
Event Type
Other
Date Received
February 21, 2013
Date of Event
January 5, 2011
Report Date
February 21, 2013
Manufacturer
ASPEN SURGICAL PRODUCTS, INC.
Product Code
GAB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BROKEN NEEDLE SAMPLES WERE SUBJECT TO A MICRO-HARDNESS TEST AND VICKER'S HARDNESS TEST. VISUALLY INSPECTED AS WELL. SAMPLES PASSED HARDNESS TESTS WITH NO FAILURE. VISUAL INSPECTION SHOWED THAT THE NEEDLE HAD EXCESSIVE GRIPPING MARKS (HEAVY CUTS INTO THE SURFACE) FROM HOLDING DEVICE (FORECEPS) ON BOTH SEGMENTS OF THE NEEDLE. SUBJECTED TO SEVERE MECHANICAL DAMAGE.

Description of Event or Problem · 1

SURGEON SUTURING SHOULDER NEEDLE BROKE OFF IN PT. SURGEON WAS ABLE TO RETRIEVE BROKEN PIECE FROM PT'S BODY. NEEDLE BROKE IN THE MIDDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
75451 RICHARD ALLEN SURGICAL NEEDLE EYED SURGICAL NEEDLE GAB ASPEN SURGICAL PRODUCTS, INC. 2015-06-03

Patients

Seq Age Sex Outcome Treatment
1 Other