RICHARD ALLEN SURGICAL NEEDLE
MDR report key: 2990035
·
Received February 21, 2013
Report
- Report Number
- 1836161-2013-00022
- Event Type
- Other
- Date Received
- February 21, 2013
- Date of Event
- January 5, 2011
- Report Date
- February 21, 2013
- Manufacturer
- ASPEN SURGICAL PRODUCTS, INC.
- Product Code
- GAB
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
BROKEN NEEDLE SAMPLES WERE SUBJECT TO A MICRO-HARDNESS TEST AND VICKER'S HARDNESS TEST. VISUALLY INSPECTED AS WELL. SAMPLES PASSED HARDNESS TESTS WITH NO FAILURE. VISUAL INSPECTION SHOWED THAT THE NEEDLE HAD EXCESSIVE GRIPPING MARKS (HEAVY CUTS INTO THE SURFACE) FROM HOLDING DEVICE (FORECEPS) ON BOTH SEGMENTS OF THE NEEDLE. SUBJECTED TO SEVERE MECHANICAL DAMAGE.
Description of Event or Problem · 1
SURGEON SUTURING SHOULDER NEEDLE BROKE OFF IN PT. SURGEON WAS ABLE TO RETRIEVE BROKEN PIECE FROM PT'S BODY. NEEDLE BROKE IN THE MIDDLE.