FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2990030 · Received March 5, 2013

Report

Report Number
2531779-2013-02374
Event Type
Malfunction
Date Received
March 5, 2013
Date of Event
February 13, 2013
Report Date
February 13, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWED LOSS OF PRIME. A REWIND, LOAD, AND PRIME SEQUENCE WAS PERFORMED WITH NO LOSS OF PRIMES OCCURRING. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO LOSS OF PRIME OCCURRING. INVESTIGATION REVEALED THAT THE FORCE SENSOR WAS OUT OF CALIBRATION. THERE WAS EVIDENCE OF CONTAMINATION OBSERVED ON THE FORCE SENSOR PLATE. UNRELATED TO THE FORCE SENSOR ISSUE, EVALUATION REVEALED A DISCOLORED/FADED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE OWNER'S BOOKLET INSTRUCTS THE USER TO CALL ANIMAS CUSTOMER SERVICE IF THE USER SUSPECTS THAT THE PRODUCT IS DAMAGED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR LOSS OF PRIME. INVESTIGATION REVEALED CONTAMINATION ON THE FORCE SENSOR. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
94268 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 65 YR