FDA Adverse Event Malfunction Summary report: N

MAXSTAR CEILING MOUNT LIGHT

MDR report key: 2989919 · Received February 28, 2013

Report

Report Number
1017522-2013-00005
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
January 30, 2013
Report Date
January 30, 2013
Manufacturer
MARUS
Product Code
EAZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE LENS HEAT SHIELD/HOLDER FROM THE DENTAL LIGHT HAS A THREE PRONG LOCKING MECHANISM WHICH LOCKS THE LENS HEAT SHIELD/COVER INTO PLACE. THE LENS HEAT SHIELD/COVER HAS TO BE REMOVED BY THE END USER WHEN CLEANING THE LENS OR REPLACING THE HALOGEN BULB DURING ROUTINE OPERATOR MAINTENANCE. THE INSTALLATION INSTRUCTIONS, USE AND CARE INSTRUCTIONS, AND LABELING CLEARLY STATE TO ENSURE THE LENS HEAT SHIELD/HOLDER IS PROPERLY SECURE BY SNAPPING THE PART BACK INTO PLACE. DENTAL EQUIPMENT LLC (DBA MARUS) INITIATED A RECALL ON JUNE 10, 2011 TO ADD A TETHER TO THE LENS HEAT SHIELD/COVER TO KEEP IT FROM FALLING OFF THE DENTAL LIGHT IF IT IS NOT RE-INSTALLED PROPERLY. THE FDA CLOSED THIS RECALL ON 10/26/2012 (PLEASE REFERENCE FDA RECALL Z-2834-2011). MARUS NOTIFIED THE DISTRIBUTOR OF THE RECALL NUMEROUS TIMES WHILE THE RECALL WAS OPEN. THE DISTRIBUTOR RESPONDED BACK TO MARUS AND ACKNOWLEDGED THE RECALL BUT APPARENTLY THE DISTRIBUTOR DID NOT INSTALL THE TETHER KIT AT THIS ONE PARTICULAR DENTAL PRACTICE. AFTER FOLLOWING BACK UP WITH THE DISTRIBUTOR, THE DISTRIBUTOR WENT TO THE DENTAL OFFICE AND INSTALLED THE TETHER KIT ON (B)(4) 2013.

Description of Event or Problem · 1

A DOCTOR WAS PERFORMING ROUTINE MEDICAL TREATMENT TO A PATIENT WHEN THE LENS HEAT SHIELD/HOLDER FROM THE DENTAL LIGHT CAME OFF THE LIGHT HEAD AND FELL DOWN ONTO THE PATIENTS FACE AND NECK AREA AND BURNED THE PATIENT. GENERAL FIRST AID ONLY WAS REQUIRED AS THE DR. APPLIED A COLD PACK TO THE RED AREA ON THE PATIENTS FACE AND NECK. THERE WERE NO SERIOUS INJURIES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85998 MAXSTAR CEILING MOUNT LIGHT DENTAL OPERATING LIGHT EAZ MARUS CL1000

Patients

Seq Age Sex Outcome Treatment
1 55 YR