ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00153
- Event Type
- Malfunction
- Date Received
- March 4, 2013
- Report Date
- February 6, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: TWO EXPIRATORY EVAQUA LIMBS, AND FOUR UNSEALED BUT UNUSED RT340 ADULT BREATHING CIRCUIT KITS WERE RETURNED TO FPH IN (B)(4) FOR INSPECTION. THE EXPIRATORY EVAQUA LIMBS WERE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS. RESULTS: NO FAULT WAS FOUND WITH THE EXPIRATORY EVAQUA LIMBS THAT WERE INCLUDED IN THE UNSEALED KITS. THE PRESSURE TEST RESULTS WERE WITHIN SPECIFICATION. THE OTHER TWO EXPIRATORY LIMBS FAILED THE PRESSURE TEST DUE TO LEAKAGE FROM THE PATIENT END CONNECTOR. THE LEAK WAS CONFIRMED WHEN EACH LIMB WAS IMMERSED IN A WATER BATH. FURTHER INSPECTION REVEALED THAT THERE WAS INSUFFICIENT GLUE PRESENT IN THE PROXIMAL CONNECTOR OF EACH EXPIRATORY LIMB. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120829. CONCLUSION: THE OBSERVED LEAK WAS CAUSED BY INSUFFICIENT GLUE BEING INJECTED INTO THE EVAQUA LIMB CONNECTOR DURING THE ASSEMBLY PROCESS, WHICH WAS NOT PICKED UP DURING PRESSURE TESTING IN THE PRODUCTION LINE. THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." (B)(4).
(B)(4). THE RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUITS ARE EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT SOME RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUITS FAILED THE VENTILATOR LEAK TEST DUE TO A HOLE AT THE END OF THE EXPIRATORY LIMBS. THIS WAS OBSERVED PRIOR TO PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT SOME RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUITS FAILED THE VENTILATOR LEAK TEST. THIS WAS OBSERVED PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 91313 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LIMITED | RT340 | 120829 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |