FDA Adverse Event Injury Summary report: N

THERASPHERE

MDR report key: 2987540 · Received February 27, 2013

Report

Report Number
8022247-2013-00002
Event Type
Injury
Date Received
February 27, 2013
Date of Event
January 4, 2013
Report Date
February 27, 2013
Manufacturer
NORDION (CANADA) INC.
Product Code
NAW
PMA / PMN Number
H980006
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE NORDION PHYSICIAN AGREES WITH THE TREATING PHYSICIAN THAT THE EVENT WAS PROBABLY CAUSED BY THE INADVERTENT DEPOSITION OF MICROSPHERES INTO THE CYSTIC ARTERY ALTHOUGH THIS WAS NOT CONFIRMED. GENERALLY THE CYSTIC ARTERY IS NOT COILED AND DEPOSITION OF MICROSPHERES DOES NOT LEAD TO CHOLECYSTITIS OR CHOLECYSTECTOMY BUT GIVEN THE ACUTE NATURE OF THE EPISODE AND THE TIMING IN RELATION TO TREATMENT WITH THERASPHERE IT IS REASONABLE TO ASSUME A CONNECTION.

Description of Event or Problem · 1

A WOMAN WITH A HISTORY OF OVARIAN CANCER DEVELOPED CHOLECYSTITIS APPROXIMATELY 3 WEEKS AFTER HER SECOND THERASPHERE TREATMENT. SHE WAS TREATED WIT IV FLUIDS AND PAIN MEDICATION AND WAS RELEASED AFTER INITIALLY RESPONDING WELL TO TREATMENT. SEVERAL WEEKS LATER, THE PATIENT'S SYMPTOMS WORSENED AND SHE WAS TREATED WITH A CHOLECYSTOTOMY TUBE ON (B)(6) 2012 BUT EVENTUALLY UNDERWENT A CHOLECYSTECTOMY ON (B)(6) 2013. SHE WAS DISCHARGED HOME BUT HER POST-OP COURSE WAS COMPLICATED BY A BILE LEAK WHICH WAS TREATED AND THE PATIENT RELEASED. THE PATIENT RETURNED TO THE HOSPITAL ON (B)(6) 2013 FOR ABDOMINAL PAIN AND NAUSEA AND WAS HYDRATED AND SENT HOME. ON (B)(6), THE PATIENT AGAIN PRESENTED TO EMERGENCY ROOM WITH ABDOMINAL PAIN AND INTRACTABLE NAUSEA AND VOMITING. IMAGING REVEALED FLUID COLLECTION AND AIR WHICH ALONG WITH AN ELEVATED WBC COUNT SUGGESTED AN ABSCESS. THE PATIENT WAS TREATED WITH IV ANTIBIOTICS AND PAIN MEDICATION UNTIL SHE WAS DISCHARGED ON (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85426 THERASPHERE YTTRIUM-90 MICROSPHERES NAW NORDION (CANADA) INC.

Patients

Seq Age Sex Outcome Treatment
1 48 YR Hospitalization| L