FDA Adverse Event
Malfunction
Summary report: N
ANODYNE THERAPY
MDR report key: 2985387
·
Received February 5, 2013
Report
- Report Number
- 1055581-2013-00001
- Event Type
- Malfunction
- Date Received
- February 5, 2013
- Date of Event
- December 21, 2012
- Report Date
- February 5, 2013
- Manufacturer
- ANODYNE THERAPY, LLC
- Product Code
- ILY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE SYSTEM INVOLVED IN THIS COMPLAINT WAS RETURNED FOR EVALUATION AND A TRANSISTOR FAILURE WAS DETECTED. NO ADVERSE EVENT WAS REPORTED, HOWEVER, THIS TYPE OF MALFUNCTION COULD POTENTIALLY CAUSE AN ADVERSE EVENT. THE NUMBER OF INCIDENTS OF THIS TYPE REMAIN WITHIN THE EXPECTED RATE FOR THIS PRODUCT BASED UPON THE NUMBER OF INCIDENTS REPORTED AND THE NUMBER OF UNITS IN DISTRIBUTION. COMPANY CONTINUES TO MONITOR AND TREND EVENTS OF THIS TYPE.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE BAR GRAPH DISPLAY DID NOT GO TO THE MAXIMUM SETTING. WE REQUESTED THAT THE UNIT BE RETURNED FOR EVALUATION, AND A TRANSISTOR WAS FOUND TO NOT BE WORKING AS INTENDED. THERE WAS NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 47503 | ANODYNE THERAPY | INFRARED LAMP | ILY | ANODYNE THERAPY, LLC | 480 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |