FDA Adverse Event
Malfunction
Summary report: N
UNIFY CRT-D, DF4 CONNECTOR
MDR report key: 2984501
·
Received February 28, 2013
Report
- Report Number
- 2938836-2013-00566
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NIK
- PMA / PMN Number
- P030054
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED. THE REPORTED NO COMMUNICATION WAS CONFIRMED AND WAS DUE TO A DEPLETED BATTERY. A LONGEVITY CALCULATION WAS PERFORMED AND WAS FOUND TO BE BELOW THE EXPECTED LIMITS. THE DEVICE WAS TESTED ON THE BENCH AND ON OUR AUTOMATED TESTING EQUIPMENT AND NO HIGH CURRENT DRAIN SOURCES WERE FOUND. THE BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER ANALYSIS. THE CAUSE OF THE BATTERY DEPLETIION COULD NOT BE DETERMINED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED. DEVICE WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88577 | UNIFY CRT-D, DF4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | CD3235-40Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |