FDA Adverse Event Malfunction Summary report: N

UNIFY CRT-D, DF4 CONNECTOR

MDR report key: 2984501 · Received February 28, 2013

Report

Report Number
2938836-2013-00566
Event Type
Malfunction
Date Received
February 28, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED. THE REPORTED NO COMMUNICATION WAS CONFIRMED AND WAS DUE TO A DEPLETED BATTERY. A LONGEVITY CALCULATION WAS PERFORMED AND WAS FOUND TO BE BELOW THE EXPECTED LIMITS. THE DEVICE WAS TESTED ON THE BENCH AND ON OUR AUTOMATED TESTING EQUIPMENT AND NO HIGH CURRENT DRAIN SOURCES WERE FOUND. THE BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER ANALYSIS. THE CAUSE OF THE BATTERY DEPLETIION COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED. DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88577 UNIFY CRT-D, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3235-40Q NA

Patients

Seq Age Sex Outcome Treatment
1