FDA Adverse Event
Malfunction
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
MDR report key: 2984492
·
Received February 28, 2013
Report
- Report Number
- 2938836-2013-00451
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- January 30, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Additional Manufacturer Narrative · 1
ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT HIGH HVLI AND NOISE WERE NOTED. WITH THE SVC VECTOR PROGRAMMED OFF, SHOCK IMPEDANCE WAS IN-RANGE. AT TIME OF EXPLANT, A LITTLE AIR BUBBLE WAS NOTED UNDER THE INSULATION BETWEEN THE RV AND SVC COILS. THE LEAD WAS REPLACED. THE PATIENT IS IN GOOD CONDITION AFTER THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88574 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7120Q/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR |