FDA Adverse Event Malfunction Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 2984492 · Received February 28, 2013

Report

Report Number
2938836-2013-00451
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
January 30, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Additional Manufacturer Narrative · 1

ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT HIGH HVLI AND NOISE WERE NOTED. WITH THE SVC VECTOR PROGRAMMED OFF, SHOCK IMPEDANCE WAS IN-RANGE. AT TIME OF EXPLANT, A LITTLE AIR BUBBLE WAS NOTED UNDER THE INSULATION BETWEEN THE RV AND SVC COILS. THE LEAD WAS REPLACED. THE PATIENT IS IN GOOD CONDITION AFTER THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88574 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120Q/65 NA

Patients

Seq Age Sex Outcome Treatment
1 49 YR