FDA Adverse Event
Malfunction
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 2984465
·
Received February 28, 2013
Report
- Report Number
- 2938836-2013-00325
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- January 17, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0457
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Additional Manufacturer Narrative · 1
EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE FOUND AT 55.0-57.2CM FROM THE CONNECTOR PIN. INTERNAL INSULATION ABRASION WAS FOUND AT 57.1-57.4CM FROM THE CONNECTOR PIN. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED IN CLINIC FOR A NORMAL FOLLOW UP. EXTERNALIZED CONDUCTORS WERE OBSERVED VIA DIAGNOSTIC IMAGING. NO ELECTRICAL ANOMALIES WERE NOTED. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88352 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1581/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |