FDA Adverse Event
Malfunction
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR
MDR report key: 2984413
·
Received February 28, 2013
Report
- Report Number
- 2938836-2013-00518
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- January 29, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED THE LEAD WAS RETURNED FOR ANALYSIS CUT IN TWO SEGMENTS. THE OUTER SHEATH WAS ABRADED AT 7.0-7.7CM AND 8.0-9.0CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE LEAD INSULATION WAS INTACT AT THESE LOCATIONS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR LEAD REVISION DUE TO DECREASED SENSING. THE PHYSICIAN MENTIONED THAT THERE SEEMED TO BE INSULATION ABRASION AT THE PROXIMAL END OF THE LEAD. LEAD WAS EXPLANTED AND REPLACED, WITHOUT PROBLEMS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 87847 | DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7122Q/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |