FDA Adverse Event Malfunction Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR

MDR report key: 2984413 · Received February 28, 2013

Report

Report Number
2938836-2013-00518
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
January 29, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED THE LEAD WAS RETURNED FOR ANALYSIS CUT IN TWO SEGMENTS. THE OUTER SHEATH WAS ABRADED AT 7.0-7.7CM AND 8.0-9.0CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE LEAD INSULATION WAS INTACT AT THESE LOCATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED FOR LEAD REVISION DUE TO DECREASED SENSING. THE PHYSICIAN MENTIONED THAT THERE SEEMED TO BE INSULATION ABRASION AT THE PROXIMAL END OF THE LEAD. LEAD WAS EXPLANTED AND REPLACED, WITHOUT PROBLEMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87847 DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7122Q/65 NA

Patients

Seq Age Sex Outcome Treatment
1