FDA Adverse Event Malfunction Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 2984383 · Received February 28, 2013

Report

Report Number
2938836-2013-00425
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
December 28, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

A FRACTURE OF THE CONDUCTOR WAS FOUND AT 3.1CM FROM THE CONNECTOR PIN CONSISTENT WITH FATIGUE. THIS FRACTURE IS CONSISTENT WITH THE OBSERVATION FROM THE FIELD OF HIGH HV LEAD IMPEDANCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT HIGH HV IMPEDANCE WAS OBSERVED. THE LEAD WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87695 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120/65 NA

Patients

Seq Age Sex Outcome Treatment
1