FDA Adverse Event
Malfunction
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 2984339
·
Received February 28, 2013
Report
- Report Number
- 2938836-2013-00282
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0457
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Additional Manufacturer Narrative · 1
EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE FOUND AT 7.0-9.1CM FROM THE LEAD TIP. INTERNAL INSULATION ABRASIONS WERE FOUND AT 25.8-27.0CM AND 27.8- 28.2CM FROM THE LEAD TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. INTERNAL INSULATION ABRASION WAS FOUND AT 28.5-29.5CM FROM THE LEAD TIP. THE ETFE COATING WAS ABRADED AT THIS LOCATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT EXTERNALIZED CONDUCTORS WERE NOTED VIA FLUOROSCOPY. ALL ELECTRICAL MEASUREMENTS WERE NORMAL. LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 86218 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1580/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|