DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR

MDR report key: 2984286 · Received February 28, 2013

Report

Report Number
2938836-2013-00444
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
January 4, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED A PARTIAL LEAD WAS RETURNED CUT IN TWO SEGMENTS. THE SEGMENTS OF THE LEAD THAT WERE RETURNED WERE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT HIGH HV LEAD IMPEDANCE WAS OBSERVED. LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85838 DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120Q/65 NA

Patients

Seq Age Sex Outcome Treatment
1 75 YR (B)(4)