FDA Adverse Event Malfunction Summary report: N

RIATA ST PASSIVE FIXATION

MDR report key: 2984270 · Received February 28, 2013

Report

Report Number
2938836-2013-00403
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
December 27, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0458
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

THE REPORTED INSULATION ABRASION OBSERVED FROM THE FIELD WAS CONFIRMED. INTERNAL INSULATION ABRASIONS WERE FOUND AT 10.0 TO 11.5CM AND 11.0 TO 12.0CM FROM THE DISTAL TIP. THE RE AND RV CONDUCTORS ETFE COATING WAS INTACT AT BOTH LOCATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT EXTERNALIZED CONDUCTORS WERE OBSERVED VIA FLUOROSCOPY. THE LEAD WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87667 RIATA ST PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7040/65 NA

Patients

Seq Age Sex Outcome Treatment
1 65 YR