FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 2984267 · Received February 28, 2013

Report

Report Number
2938836-2013-00369
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
December 20, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED A COMPLETE LEAD WAS RETURNED CUT IN TWO SEGMENTS FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS FOUND AT 7.1-7.6CM, 7.8-8.4CM, AND 26.0-26.3CM FROM THE HELIX END, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE. THE ETFE COATING WAS INTACT AT THE SAME LOCATION. (B)(4) - NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87675 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7000/65 NA

Patients

Seq Age Sex Outcome Treatment
1