FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 2984248 · Received February 28, 2013

Report

Report Number
2938836-2013-00367
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
December 5, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. RELIABILITY LABORATORY TECHNICIAN; (B)(6). FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASIONS WERE FOUND AT 14.0-14.3CM AND 17.0-17.3 CM FROM THE PROXIMAL END OF THE SVC SHOCK COIL, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.