FDA Adverse Event
Malfunction
Summary report: N
RIATA ST ACTIVE FIXATION
MDR report key: 2984248
·
Received February 28, 2013
Report
- Report Number
- 2938836-2013-00367
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- December 5, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. RELIABILITY LABORATORY TECHNICIAN; (B)(6). FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASIONS WERE FOUND AT 14.0-14.3CM AND 17.0-17.3 CM FROM THE PROXIMAL END OF THE SVC SHOCK COIL, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 86767 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7000/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |