RIATA PASSIVE FIXATION
Report
- Report Number
- 2938836-2013-00247
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- December 3, 2012
- Report Date
- March 12, 2018
- Manufacturer
- ST. JUDE MEDICAL, INC.
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0457
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
.
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
(B)(4).
NEW INFORMATION RECEIVED NOTES THAT A VARIATION IN SENSING AND PACING THRESHOLD WAS OBSERVED. THE NOISE WAS NOT REPRODUCED WITH PROVOCATIVE TESTING AND POCKET MANIPULATION. THE LEAD REMAINS IMPLANTED AND PATIENT WILL BE MONITORED. PATIENT MEDICAL CONDITION CONTINUES TO BE GOOD.
IT WAS REPORTED DURING DEVICE CHANGE OUT, EXTERNALIZED CONDUCTORS WERE OBSERVED VIA FLUOROSCOPY. NO ELECTRICAL ANOMALIES WERE NOTED. LEAD REMAINS IMPLANTED. PATIENT CONDITION AFTER THE EVENT WAS GOOD.
NEW INFORMATION RECEIVED NOTES THAT WHEN THE DEVICE WAS INTERROGATED, A CAPTURE ANOMALY AND NOISE WERE OBSERVED. THE NOISE WAS UNABLE TO BE REPRODUCED IN REAL TIME. THE PHYSICIAN ELECTED NOT TO TAKE ANY ACTION. THE PATIENT CONDITION WAS GOOD.
NEW INFORMATION RECEIVED NOTES THAT THE PROGRAMMING CHANGES WERE MADE. THE PATIENT CONDITION WAS GOOD.
NEW INFORMATION WAS RECEIVED THAT THE PACING THRESHOLD ON THE RIGHT VENTRICULAR CHANNEL WAS VARIABLE. ALL OTHER ELECTRICAL VALUES WERE STABLE AND IN RANGE. NO INTERVENTION WAS PERFORMED AND THE PATIENT WAS STABLE AND WILL CONTINUE TO BE MONITORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 86523 | RIATA PASSIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC. | 1571/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |