FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 2984137 · Received February 28, 2013

Report

Report Number
3004209178-2013-91470
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
January 17, 2013
Report Date
February 26, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSPECTED THREE OPENED AND USED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER SPECIFICATION. NO LEAKAGE ANOMALIES WERE OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED PAST THE O-RINGS AND INTO THE RESERVOIR COMPARTMENT. THE BLOOD GLUCOSE READING WAS 21.4MMOL/L. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85986 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A H8418510

Patients

Seq Age Sex Outcome Treatment
1