FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 2984136 · Received February 28, 2013

Report

Report Number
3004209178-2013-91464
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 1, 2013
Report Date
February 20, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

INSPECTED FOUR OPENED AND USED RESERVOIRS. PERFORMED PRE-FILL RESERVOIRS TEST. RESERVOIRS AND TRANSFER GUARDS PASSED PER INSPECTION. FOUND NO LEAKAGE ANOMALIES DURING FILLING. NO PHYSICAL OR COSMETIC ANOMALIES DURING TEST. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER SPECIFICATION. NO LEAKAGE ANOMALIES WERE OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. RESERVOIRS CONNECTED AND LOCKED IN PLACE PROPERLY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN LEAKED AT THE SNAP CAP. THE BLOOD GLUCOSE READING WAS 200MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88175 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A H7887608

Patients

Seq Age Sex Outcome Treatment
1 42 YR