FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2984123
·
Received February 28, 2013
Report
- Report Number
- 3004209178-2013-91506
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- February 10, 2013
- Report Date
- February 10, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH CONSTANT TONE AND FROZEN SCREEN AS A RESULT OF A FAULTY COMPONENT ON THE MOTHERBOARD. UNABLE TO CONFIRM THE BUTTON/KEYPAD ANOMALY DUE TO THE FROZEN SCREEN.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD A CONSTANT TONE. THE BLOOD GLUCOSE READING WAS 312MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CALLER STATED THAT THE BUTTONS WERE UNRESPONSIVE. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 85992 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |