FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2984102 · Received February 28, 2013

Report

Report Number
3004209178-2013-91473
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 13, 2013
Report Date
February 26, 2013
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST FOLLOWED BY A MOTOR ERROR ALARM DURING REWIND AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE. UNABLE TO PERFORM THE FUNCTIONAL TEST DUE TO MOTOR ANOMALY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING WAS 177MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CALLER STATED THAT THE DEVICE WAS EXPOSED TO A HIGH MAGNETIC FIELD WHILE HAVING AN MRI. ASSISTED THE CUSTOMER TO RUN THE DISPLACEMENT AND PASSED. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88083 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 56 YR