FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2984102
·
Received February 28, 2013
Report
- Report Number
- 3004209178-2013-91473
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- February 13, 2013
- Report Date
- February 26, 2013
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST FOLLOWED BY A MOTOR ERROR ALARM DURING REWIND AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE. UNABLE TO PERFORM THE FUNCTIONAL TEST DUE TO MOTOR ANOMALY.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING WAS 177MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CALLER STATED THAT THE DEVICE WAS EXPOSED TO A HIGH MAGNETIC FIELD WHILE HAVING AN MRI. ASSISTED THE CUSTOMER TO RUN THE DISPLACEMENT AND PASSED. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88083 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MMT-523NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |