FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2984065 · Received February 28, 2013

Report

Report Number
3004209178-2013-91431
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 15, 2013
Report Date
February 15, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO MOTOR ANOMALY. THE INSULIN PUMP RECEIVED WITH STRIPPED BATTERY CAP COIN SLOT.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING BOLUS DELIVERY. TROUBLESHOOTING WAS PERFORMED. THE BLOOD GLUCOSE READING WAS 212MG/DL. THE CALLER STATED THAT DRIVE SUPPORT CAP WAS FLUSH. THE CUSTOMER STATED THAT THE DEVICE WAS EXPOSED TO AN MRI. THE CUSTOMER WAS IN THE HOSPITAL FOR NON DIABETES RELATED ISSUES. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86463 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 55 YR