PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2984028 · Received February 28, 2013

Report

Report Number
2032227-2013-00883
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 17, 2013
Report Date
March 19, 2018
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Additional Manufacturer Narrative · 1

PUMP ALARMED MOTOR ERROR DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST AND EXCESSIVE NO DELIVERY TEST DUE TO MOTOR ERROR ALARMS. PUMP RECEIVED WITH MINOR SCRATCHED LCD WINDOW, BLEEDING, CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED LCD WINDOW, CRACKED BELT CLIP SLOT, CRACKED BATTERY TUBE THREADS, STAINED END CAP STICKER, STAINED ADDRESS/SERIAL NUMBER LABEL AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. NOTHING FURTHER WAS REPORTED.