TALENT ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2013-00359
- Event Type
- Injury
- Date Received
- February 28, 2013
- Date of Event
- March 8, 2013
- Report Date
- March 8, 2013
- Manufacturer
- MEDTRONIC CARDIOVASCULAR SANTA ROSA
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). METHOD: (FILM). RESULTS: INHERENT RISK OF PROCEDURE (STENT GRAFT MIGRATION, STENT FRACTURE); PATIENT'S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (DISEASE PROGRESSION, AORTIC NECK DILATATION AND ANGULATION). CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (DISEASE PROGRESSION, AORTIC NECK DILATATION AND ANGULATION).
ADDITIONAL INFORMATION WAS RECEIVED AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT'S AORTIC NECK MEASURED LESS THAN 10 MM IN LENGTH AND THE OTHER SIDE DID NOT HAVE VERY MUCH PURCHASE. THEREFORE THE PHYSICIAN ELECTED TO COMPROMISE THE RENAL ARTERY. TWO ENDURANT AORTIC CUFFS WERE IMPLANTED AND THERE WAS 50% ENCROACHMENT ON THE LEFT RENAL ARTERY; HOWEVER, THERE WAS GOOD BLOOD FLOW THROUGH THE RENAL ARTERY UPON COMPLETION OF THE PROCEDURE. THE PROXIMAL TYPE I ENDOLEAK WAS SUCCESSFULLY RESOLVED.
A TALENT STENT GRAFT WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 56 MM DIAMETER ABDOMINAL AORTIC ANEURYSM APPROXIMATELY FOUR YEARS AGO. THE AORTIC NECK WAS 27-28 MM IN DIAMETER AND DESCRIBED AS SHORT AND ANGULATED AT THE TIME OF IMPLANT. CURRENTLY THE AORTIC NECK DIAMETER IS 32-33MM, AND THE FEMORAL ARTERIES ARE 7.5MM IN DIAMETER BILATERALLY. IT WAS REPORTED THAT THE TALENT BIFURCATED STENT GRAFT HAD MIGRATED APPROXIMATELY 4 CM. THE MIGRATION WAS ALSO NOTED TWO YEARS AND ONE YEAR AGO. THE RADIOLOGIST NOTED NO EVIDENCE OF AN ENDOLEAK AND THE GRAFT IS IN THE ANEURYSM SURROUNDED BY THROMBUS. INTERVENTION HAS NOT BEEN PERFORMED AND THE PATIENT WILL BE MONITORED BY THE PHYSICIAN. NO FURTHER INFORMATION WAS PROVIDED. REVIEW OF RETURNED FILMS POST-IMPLANT SHOWED THAT THE BIFURCATED STENT GRAFT WAS 4-5CM BELOW THE RENAL ARTERIES IN THE ANEURYSM SAC. THE IPSILATERAL LIMB WAS ON THE LEFT SIDE; AN ANEURX LIMB EXTENDED INTO THE LEFT EXTERNAL ILIAC. THE ANEURX CONTRALATERAL AND CONTRALATERAL EXTENSION WERE PLACED ON THE RIGHT SIDE. THE CONTRAST APPEARED TO BE CONTAINED WITHIN THE THROMBUS; HOWEVER, IT IS UNKNOWN IF THE SAC MAY STILL BE PRESSURIZED. THE AAA DIAMETER MEASURED 6-7CM. THE PROXIMAL NECK DIAMETER AT THE RENAL ARTERIES MEASURED 28 X 34MM AND THE AORTIC NECK WAS ANGULATED 70DEG L-R AND A-P. EARLIER FOLLOW-UP IMAGES WERE NOT PROVIDED. THE CAUSE OF THE MIGRATION APPEARS TO BE DISEASE PROGRESSION WITH NECK DILATATION AND ANGULATED NECK. ALSO OBSERVED WAS A 2CM LENGTH STRAIGHT SEGMENT OF THE BIFURCATE BARE SPRING WHICH HAD FRACTURED AND SEPARATED; IT WAS POSITIONED APPROXIMATELY 2CM BELOW THE RENALS ON THE LEFT SIDE OF THE PROXIMAL NECK/AAA. THE CAUSE OF THE FRACTURE IS UNCERTAIN; IT IS POSSIBLE THAT DISEASE PROGRESSION (ANGULATED NECK) AND MORPHOLOGY CHANGES MAY HAVE CONTRIBUTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 86116 | TALENT ABDOMINAL STENT GRAFT SYSTEM | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC CARDIOVASCULAR SANTA ROSA | V00191537 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00071 YR | Required Intervention |