FDA Adverse Event Malfunction Summary report: N

TENDRIL ST

MDR report key: 2983972 · Received February 28, 2013

Report

Report Number
2017865-2013-01498
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
January 15, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED. (B)(6). NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THE OUTER INSULATION WAS DEGRADED AT 4.5 CM FROM THE CONNECTOR PIN.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87475 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1