FDA Adverse Event Injury Summary report: N

TENDRIL SDX

MDR report key: 2983919 · Received February 28, 2013

Report

Report Number
2017865-2013-01463
Event Type
Injury
Date Received
February 28, 2013
Date of Event
January 16, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED LOW IMPEDANCE, HIGH THRESHOLDS DUE TO CLAVICULAR CRUSH DAMAGE. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88600 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1688TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention