FDA Adverse Event Injury Summary report: N

ISOFLEX S

MDR report key: 2983879 · Received February 28, 2013

Report

Report Number
2017865-2013-01431
Event Type
Injury
Date Received
February 28, 2013
Date of Event
January 7, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

FINAL ANALYSIS FOUND THE LEAD WAS RETURNED IN THREE PIECES. THE PROXIMAL COIL WAS EXPOSED DUE TO CLAVICULAR CRUSH LOCATED AT 30.8 CM FROM THE ELECTRODE TIP.

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87457 ISOFLEX S PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1646T/58 NA

Patients

Seq Age Sex Outcome Treatment
1 25 YR Required Intervention