FDA Adverse Event Malfunction Summary report: N

QUICKFLEX XL LV LEAD

MDR report key: 2983850 · Received February 28, 2013

Report

Report Number
2017865-2013-01357
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
January 24, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
Z1483
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RELIABILITY LABORATORY TECHNICIAN, (B)(4). NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND AN ABRASION AT 3.7 CM TO 4.0 CM AND AT 5.8 CM TO 6.1 CM FROM THE DISTAL TIP ELECTRODE. THE INNER COIL WAS EXPOSED AT 5.7 CM FROM THE DISTAL TIP ELECTRODE. ABRASIONS ARE CONSISTENT WITH CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88694 QUICKFLEX XL LV LEAD PERMANENT PACEMAKER ELECTRODE NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1158T/86 NA