FDA Adverse Event
Malfunction
Summary report: N
QUICKFLEX XL LV LEAD
MDR report key: 2983850
·
Received February 28, 2013
Report
- Report Number
- 2017865-2013-01357
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- January 24, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NIK
- PMA / PMN Number
- P030054
- Removal / Correction Number
- Z1483
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
RELIABILITY LABORATORY TECHNICIAN, (B)(4). NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND AN ABRASION AT 3.7 CM TO 4.0 CM AND AT 5.8 CM TO 6.1 CM FROM THE DISTAL TIP ELECTRODE. THE INNER COIL WAS EXPOSED AT 5.7 CM FROM THE DISTAL TIP ELECTRODE. ABRASIONS ARE CONSISTENT WITH CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88694 | QUICKFLEX XL LV LEAD | PERMANENT PACEMAKER ELECTRODE | NIK | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1158T/86 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |