TENDRIL STS
MDR report key: 2983848
·
Received February 28, 2013
Report
- Report Number
- 2017865-2013-01555
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- January 9, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DTB
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. RELIABILITY LABORATORY TECHNICIAN, (B)(4). NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT THE OUTER INSULATION WAS ABRADED AT 11.0 CM TO 11.5 CM FROM THE CONNECTOR PIN. ABRASIONS ARE INDICATIVE OF EXPOSURE TO CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.