FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2983728 · Received February 28, 2013

Report

Report Number
2955842-2013-00671
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
December 17, 2012
Report Date
January 29, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS FOUND WITH A BROKEN PITCH CABLE AT THE WRIST. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CABLE CRIMP REMAINED IN THE CLEVIS. ENGINEERING CONCLUDED THAT THE NON-INTUITIVE MOTION WAS PROBABLY CAUSED BY THE BROKEN PITCH CABLE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DAVINCI S SURGICAL PROCEDURE, THE CUSTOMER REPORTED THAT THE FENESTRATED BIPOLAR INSTRUMENT WAS UNABLE TO ROTATE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87628 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10120905 347

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES