FENESTRATED BIPOLAR FORCEPS INSTRUMENT
MDR report key: 2983728
·
Received February 28, 2013
Report
- Report Number
- 2955842-2013-00671
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- December 17, 2012
- Report Date
- January 29, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS FOUND WITH A BROKEN PITCH CABLE AT THE WRIST. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CABLE CRIMP REMAINED IN THE CLEVIS. ENGINEERING CONCLUDED THAT THE NON-INTUITIVE MOTION WAS PROBABLY CAUSED BY THE BROKEN PITCH CABLE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A DAVINCI S SURGICAL PROCEDURE, THE CUSTOMER REPORTED THAT THE FENESTRATED BIPOLAR INSTRUMENT WAS UNABLE TO ROTATE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 87628 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M10120905 347 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES |