FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2983721 · Received February 28, 2013

Report

Report Number
1416980-2013-04989
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 20, 2013
Report Date
February 20, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED. THE LOT NUMBER IS UNKNOWN; THEREFORE, A BATCH REVIEW WILL NOT BE CONDUCTED. THIS COMPLAINT FOR SE 2240 IS NOT CONFIRMED BECAUSE THE DISPOSABLE SET WAS NOT RETURNED TO BAXTER FOR EVALUATION, THE LOT NUMBER WAS NOT PROVIDED FOR A BATCH REVIEW AND THE USER DID NOT DESCRIBE ANY TYPES OF USE ERRORS OR OTHER ISSUES THAT MIGHT HAVE CAUSED THE REPORTED EVENT. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REPORTING A SYSTEM ERROR (SE) 2240/2367 (AIR IN SET) DURING DWELL 3 OF 4 ON THE HOME CHOICE (HC). THE TECHNICAL SERVICE REPRESENTATIVE (TSR) INSTRUCTED THE HOME PATIENT (HP) TO CYCLE THE POWER THROUGH THE SYSTEM ERROR 2367 THEN TO PRESS GO TO START . THE HP WAS NOT SURE WHAT TO DO NEXT. THE TSR HAD THE HP DISCONNECT AND REMOVE THE CASSETTE. THE TSR EXPLAINED THE ALARM AND ADVISED THE HP TO CONTACT THE NURSE. THERE WAS PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87965 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 55 YR HOMECHOICE