FDA Adverse Event
Malfunction
Summary report: N
TENDRIL SDX
MDR report key: 2983716
·
Received February 28, 2013
Report
- Report Number
- 2017865-2013-01408
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- December 5, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVN
- PMA / PMN Number
- P960013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED RELIABILITY LABORATORY TECHNICIAN, (B)(4). NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT THE INSULATION WAS DAMAGED AT 22.2 CM TO 23.0 CM FROM THE CONNECTOR PIN DUE TO CLAVICULAR CRUSH.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 87541 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | NVN | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1488TC/46 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |