FDA Adverse Event Malfunction Summary report: N

TENDRIL DX

MDR report key: 2983714 · Received February 28, 2013

Report

Report Number
2017865-2013-01378
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
December 15, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(6). NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT THE INSULATION WAS ABRADED AT 5.3 CM TO 5.5 CM FROM THE CONNECTOR END. THE ABRASION WAS CONSISTENT WITH CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86037 TENDRIL DX PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1388TC/58 NA

Patients

Seq Age Sex Outcome Treatment
1