FDA Adverse Event
Injury
Summary report: N
TENDRIL SDX
MDR report key: 2983694
·
Received February 28, 2013
Report
- Report Number
- 2017865-2013-01469
- Event Type
- Injury
- Date Received
- February 28, 2013
- Date of Event
- October 11, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVN
- PMA / PMN Number
- P960013
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
FINAL ANALYSIS FOUND THAT THE INSULATION WAS ABRADED AT 5.8 CM - 6.2 CM FROM THE HELIX END, DUE TO FRICTION WITH ANOTHER DEVICE.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED LOSS OF SENSING AND CAPTURE IN BIPOLAR MODE. WHEN PROGRAMMED TO UNIPOLAR MODE THE LEAD EXHIBITED HIGH THRESHOLDS. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88020 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | NVN | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1688TC/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Required Intervention |