FDA Adverse Event Injury Summary report: N

TENDRIL SDX

MDR report key: 2983694 · Received February 28, 2013

Report

Report Number
2017865-2013-01469
Event Type
Injury
Date Received
February 28, 2013
Date of Event
October 11, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

FINAL ANALYSIS FOUND THAT THE INSULATION WAS ABRADED AT 5.8 CM - 6.2 CM FROM THE HELIX END, DUE TO FRICTION WITH ANOTHER DEVICE.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED LOSS OF SENSING AND CAPTURE IN BIPOLAR MODE. WHEN PROGRAMMED TO UNIPOLAR MODE THE LEAD EXHIBITED HIGH THRESHOLDS. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88020 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1688TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention