FDA Adverse Event Injury Summary report: N

TENDRIL SDX

MDR report key: 2983685 · Received February 28, 2013

Report

Report Number
2017865-2013-01447
Event Type
Injury
Date Received
February 28, 2013
Date of Event
November 13, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

FINAL ANALYSIS FOUND THAT THE PROXIMAL INSULATION WAS ABRADED AT 7.1 - 7.4 CM FROM THE CONNECTOR PIN. THE PROXIMAL COIL WAS EXPOSED IN THE SAME AREA WHICH COULD HAVE CONTRIBUTED TO THE NOISE PROBLEM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD EXHIBITED AN INSULATION ANOMALY AND NOISE. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85913 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE NVN ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1688TC/46 NA

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention