FDA Adverse Event
Injury
Summary report: N
TENDRIL SDX
MDR report key: 2983685
·
Received February 28, 2013
Report
- Report Number
- 2017865-2013-01447
- Event Type
- Injury
- Date Received
- February 28, 2013
- Date of Event
- November 13, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVN
- PMA / PMN Number
- P960013
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Additional Manufacturer Narrative · 1
FINAL ANALYSIS FOUND THAT THE PROXIMAL INSULATION WAS ABRADED AT 7.1 - 7.4 CM FROM THE CONNECTOR PIN. THE PROXIMAL COIL WAS EXPOSED IN THE SAME AREA WHICH COULD HAVE CONTRIBUTED TO THE NOISE PROBLEM.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD EXHIBITED AN INSULATION ANOMALY AND NOISE. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 85913 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | NVN | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1688TC/46 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Required Intervention |