FDA Adverse Event Injury Summary report: N

ICL (IMPLANTABLE COLLAMER LENS)

MDR report key: 2983671 · Received February 28, 2013

Report

Report Number
2023826-2013-00173
Event Type
Injury
Date Received
February 28, 2013
Date of Event
November 10, 2012
Report Date
February 7, 2013
Manufacturer
STAAR SURGICAL COMPANY
Product Code
MTA
PMA / PMN Number
P030016
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

EVALUATION METHOD: WORK ORDER SEARCH, MEDICAL REVIEW. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND NO VISIBLE DAMAGE TO THE LENS. THE LENS WAS RETURNED DRY. THE LENS WAS REHYDRATED IN BSS FOR RE-MEASUREMENT. THE LENS LENGTH WAS MEASURED AND THE RESULT OF THE MEASUREMENT WAS COMPARED AGAINST THE ORIGINAL VALUE AND THE LENS WAS FOUND TO BE IN SPECIFICATION. MEDICAL REVIEW - THE MOST COMMON REASONS FOR EARLY POSTOPERATIVE ELEVATED IOP AFTER ICL IMPLANTATION IN THE PRESENCE OF OPTIMAL VAULTING INCLUDE : NON PATENT/FUNCTIONING IRIDECTOMIES (TOO SMALL, TOO PERIPHERAL AND/OR NOT FULLY PERMEABLE, BLOCKED BY VISCO), REMAINING VISCOELASTIC IN THE POSTERIOR CHAMBER, OVERSIZED ICL WITH ANGLE CLOSURE, TOO NARROW ANGLES/CROWDED AC DUE TO SMALL EYE ANATOMY PREOPERATIVELY, UNEXPECTED ABNORMAL ANATOMY OR TISSUE ABNORMALITIES (PRESENCE OF IRIS/CILIARY BODY CYSTS), MALPOSITIONED FOOTPLATES, ETC. CONCLUSIONS - (NO CONCLUSION CAN BE DRAWN): BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH, MEDICAL REVIEW AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).

Description of Event or Problem · 1

THE REPORTER INDICATED THE SURGEON IMPLANTED A 12.0MM ICM120V4 IMPLANTABLE COLLAMER LENS IN THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2012. THE LENS WAS EXPLANTED ON (B)(6) 2012 DUE TO ELEVATED IOP AND BLURRED VISION. THE PUPIL WAS DILATED. THE PUPIL AND IOP WERE NORMAL AFTER ICL WAS EXPLANTED. THE PATIENT'S BCVA WAS 20/25.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87755 ICL (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS MTA STAAR SURGICAL COMPANY ICM120V4 NA

Patients

Seq Age Sex Outcome Treatment
1 30 YR Required Intervention CARTRIDGE MODEL AND LOT NUMBER UNK| INJECTOR MODEL AND LOT NUMBER UNK