FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2983630 · Received February 28, 2013

Report

Report Number
1416980-2013-04974
Event Type
Malfunction
Date Received
February 28, 2013
Report Date
February 15, 2013
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE EXACT DATE THE EVENT OCCURED IS UNKNOWN. THE SAMPLE WAS NOT RETURNED TO BAXTER FOR EVALUATION. THEREFORE, THE REPORTED CONDITION COULD NOT BE CONFIRMED AND THE ROOT CAUSE COULD NOT BE DETERMINED. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INTERMATE HAD A RESERVOIR RUPTURE DURING THE END OF AN INFUSION. THE CONCOMITANT MEDICAL PRODUCTS ARE CURRENTLY UNKNOWN. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86581 INTERMATE PUMP, INFUSION, ELASTOMERIC FRN BAXTER HEALTHCARE - IRVINE 12B034

Patients

Seq Age Sex Outcome Treatment
1