FDA Adverse Event Malfunction Summary report: N

COULTER® AC T 5DIFF CP ANALYZER

MDR report key: 2983600 · Received February 28, 2013

Report

Report Number
1061932-2013-00270
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 4, 2013
Report Date
February 4, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K003677
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE FOUND THE REAGENT SYRINGE WAS CRACKED AND LEAKING LYSE REAGENT. THE LYSE REAGENT LEAKED INTO THE DRIP TRAY UNDER THE SYRINGES AND EVENTUALLY TO THE BOTTOM OF THE INSTRUMENT. THE REAGENT SYRINGE MAY CONTAIN LYSE WHILE IN OPERATION AND CLEANER WHILE IN SHUTDOWN. THE FSE REPLACED THE REAGENT SYRINGE AND RESOLVED THE LEAK. PER LABELING, BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. FAILURE MODE OF THIS EVENT WAS REAGENT SYRINGE CRACKED AND LEAKING LYSE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) A LEAK OF CLEAR SOAPY FLUID UNDERNEATH THE COULTER AC T 5DIFF CP ANALYZER. THE VOLUME OF THE LEAK WAS REPORTED BY THE CUSTOMER AS 2 TO 5 ML. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS REVIEWED. THERE IS AN EXPOSURE CONTROL PLAN AT THE FACILITY. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE), CONSISTING OF GLOVES AND A LABORATORY COAT. NO ONE WAS SPLASHED, SPRAYED OR INJURED. THERE WAS NO CONTACT TO OPEN OR BROKEN SKIN OR MUCOUS MEMBRANES. NO ONE SOUGHT MEDICAL ATTENTION. NO PATIENT RESULTS WERE AFFECTED. THERE WAS NO DEATH, INJURY OR AFFECT TO PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85659 COULTER® AC T 5DIFF CP ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER ACT 5DIFF CP NA

Patients

Seq Age Sex Outcome Treatment
1