FDA Adverse Event Death Summary report: N

SKIN-PREP WIPES BOX

MDR report key: 2983584 · Received February 28, 2013

Report

Report Number
3006760724-2013-00002
Event Type
Death
Date Received
February 28, 2013
Date of Event
February 9, 2011
Report Date
February 28, 2013
Manufacturer
SMITH & NEPHEW WOUND MANAGEMENT
Product Code
NEC
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

SAMPLES HAVE NOT BEEN RETURNED TO SMITH & NEPHEW FOR EVALUATION. INVESTIGATION INTO RETAIN BATCH RECORDS SHOW THAT INDEPENDENT MICROBIOLOGICAL TESTS WERE CONDUCTED ON THE RETAINS FOR THE LOT REPORTED. THERE WAS NO EVIDENCE OF MICROBIOLOGICAL CONTAMINATION FOUND IN THE RETAIN SAMPLES. AN ACTIVE INVESTIGATION IS CURRENTLY IN PROGRESS AND THE RESULTS OF OUR INVESTIGATION WILL BE DETAILED IN A SUPPLEMENT REPORT.

Additional Manufacturer Narrative · 1

SMITH & NEPHEW HAS COMPLETED THEIR INVESTIGATION INTO THE REPORTED INCIDENT. AFTER A THOROUGH INVESTIGATION WE WERE UNABLE TO CONFIRM ANY CAUSAL RELATIONSHIP, OR DEVICE FAILURE AS HAVING CAUSED OR CONTRIBUTED TO THE EVENT REPORTED. AN INDEPENDENT MEDICAL REVIEW OF SIMILAR COMPLAINTS CONFIRMED THAT SKIN PREP WIPES IS NOT THE ROOT CAUSE OF THE ISSUE. THERE WAS NO PRODUCT WAS RETURNED BY THE CUSTOMER SO NO TESTING COULD BE PERFORMED ON CUSTOMER SAMPLES. HOWEVER, THE RETAIN SAMPLES OF THIS LOT WERE ANALYZED BY AN INDEPENDENT ANALYTICAL LAB AND NO MICROBIAL CONTAMINATION OR GROWTH WAS NOTED. SKIN PREP WIPES IS AN ISOPROPYL ALCOHOL BASED LIQUID FILM-FORMING SKIN PREPARATION THAT, UPON APPLICATION TO INTACT SKIN, FORMS A PROTECTIVE FILM TO HELP REDUCE FRICTION DURING REMOVAL OF TAPES AND FILMS. PER THE PRODUCT LABELING, IT IS VERY IMPORTANT TO ALLOW TO DRY THOROUGHLY BEFORE ANY DEVICE IS APPLIED. WE WERE UNABLE TO DETERMINE A SPECIFIC ROOT CAUSE FOR THIS ISSUE. GIVEN THAT PRODUCT HAS SUFFICIENT WARNINGS AND THAT IT HAS BEEN RELOCATED, NO FURTHER INVESTIGATION, OR CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME.

Description of Event or Problem · 1

SKIN PREP WIPES THOUGHT TO BE CONTAMINATED WITH BACILLUS CEREUS. AFTER USING THE WIPES, THE PATIENT EXPERIENCED A BACTERIAL INFECTION WHICH RESULTED IN DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86334 SKIN-PREP WIPES BOX BANDAGE, LIQUID, SKIN PROTECTANT NEC SMITH & NEPHEW WOUND MANAGEMENT 420400 0L165

Patients

Seq Age Sex Outcome Treatment
1 Death