HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2013-04968
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- February 4, 2013
- Report Date
- February 4, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THIS REPORT IS THE SAME PATIENT AS (B)(4). THE REPORTED ISSUE WAS NOT CONFIRMED AND THE CAUSE WAS UNDETERMINED BECAUSE THE SAMPLE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE
(B)(4). THE PROBLEM WAS CONFIRMED FOR THIS USE ERROR - FAILED TO FOLLOW THERAPY STEPS COMPLAINT, BECAUSE THERE WAS A DEFINED USE ERROR. PER BAXTER LABELING, USERS ARE INSTRUCTED TO USE ASEPTIC TECHNIQUE WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. THE CAUSE IS UNDETERMINED. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
IT WAS REPORTED THE PATIENT WAS SUPPOSED TO USE A HOMECHOICE AUTOMATED PD SET WITH 3-PRONG CASSETTE, BUT INSTEAD USED A HOMECHOICE AUTOMATED PD SET WITH 4-PRONG CASSETTE TO PERFORM PERITONEAL DIALYSIS (PD) THERAPY. THE PATIENT CUT THE TUBING ON THE CASSETTE AND TAPED THE TUBING TOGETHER AND CONTINUED TO USE THIS CASSETTE FOR PD THERAPY. THE HOME CARE SERVICE REPRESENTATIVE (HCSR) INSTRUCTED THE PATIENT TO DISCONTINUE USE OF THE INCORRECT/CUT CASSETTE AND NOTIFIED THE PD NURSE OF THE ISSUE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. ADDITIONAL INFORMATION WAS REQUESTED, BUT WAS NOT AVAILABLE AT THIS TIME. THIS IS REPORT 1 OF 2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88238 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |