FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2983364 · Received February 28, 2013

Report

Report Number
1416980-2013-04951
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 1, 2013
Report Date
February 8, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF FAILURE CODE 94 WAS CONFIRMED DURING DEVICE EVALUATION. THE ASSIGNABLE CAUSE WAS IDENTIFIED AS A DAMAGED MAIN BATTERY. THE MAIN BATTERY WAS REPLACED TO CORRECT THE REPORTED CONDITION. ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON-SITE AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPORTED A FLOGARD INFUSION PUMP THAT PRESENTED "F94". IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT/USER INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.