FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2983320 · Received January 7, 2013

Report

Report Number
1720753-2013-00253
Event Type
Malfunction
Date Received
January 7, 2013
Date of Event
December 20, 2012
Report Date
January 7, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND CLEANED AND LUBRICATED THE CANDLESTICK HIGH VOLTAGE CONNECTORS. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM MADE A POPPING NOISE AND SHUT DOWN ON ITS OWN. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
7580 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1