FDA Adverse Event Malfunction Summary report: N

UNIVERSAL HIGH TORQUE DRILL

MDR report key: 2983064 · Received February 28, 2013

Report

Report Number
0001811755-2013-00400
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
February 4, 2013
Report Date
February 4, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL WAS RECEIVED BY THE MANUFACTURER AND A DEVICE EVALUATION WAS CONDUCTED. THE EVALUATION REVEALED THAT THE ROTOR WAS DAMAGED AND THE BEARINGS WERE CONTAMINATED WITH DEBRIS. THE ROTOR ASSEMBLY AND BEARINGS WERE REPLACED, AND THE DRILL WAS RETURNED TO THE USER FACILITY.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER AND AN INVESTIGATION IS ANTICIPATED. A FOLLOW UP REPORT WILL BE SUBMITTED IF THE INVESTIGATION RESULTS REQUIRE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER SALES REPRESENTATIVE, THE DRILL OPERATED WHILE IN THE LOAD MODE. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER SALES REPRESENTATIVE, THE DRILL OPERATED WHILE IN THE LOAD MODE. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87678 UNIVERSAL HIGH TORQUE DRILL DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1