FDA Adverse Event Malfunction Summary report: N

PWRD ECH FLEX 60MM

MDR report key: 2982946 · Received February 28, 2013

Report

Report Number
3005075853-2013-00920
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
January 30, 2013
Report Date
February 11, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K110385
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INCORRECT CARTRIDGE SIZE. THE ANALYSIS FOUND THAT ONE PLE60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH THE MANUAL OVERRIDE DOOR OUT OF POSITION; THE OVERRIDE LEVER WAS UP WHICH DENOTES THAT THE KNIFE WAS MANUALLY RETURNED TO HOME POSITION. AN ECR45G CARTRIDGE RELOAD WAS LOADED IN THE DEVICE. THE CARTRIDGE WAS RECEIVED UNFIRED. IT SHOULD BE NOTED THAT A 60MM DEVICE IS DESIGNED TO WORK ONLY WITH 60MM CARTRIDGES, FOR FURTHER LOADING INSTRUCTIONS PLEASE REFER TO THE ECHELON FLEX 60 POWERED IFU. PLEASE NOTE THAT IF THE REVERSE SWITCH DOES NOT RETURN THE KNIFE TO HOME POSITION THE JAWS WILL NOT OPEN. ENSURE THAT THE BATTERY PACK IS SECURELY INSTALLED AND THE INSTRUMENT HAS POWER AND THEN, TRY THE KNIFE REVERSE SWITCH AGAIN. IF THE KNIFE DOES NOT RETURN, USE THE MANUAL OVERRIDE. AFTER THE MANUAL OVERRIDE SYSTEM IS USED THE INSTRUMENT IS DISABLED AND CANNOT BE USED FOR ANY SUBSEQUENCE FIRINGS. IN ADDITION, IT SHOULD BE NOTED THAT IF THE INSTRUMENT IS PARTIALLY FIRED, SLIDE THE KNIFE REVERSE SWITCH FORWARD TO RETURN THE KNIFE TO HOME POSITION. TO OPEN THE JAWS, SQUEEZE THE CLOSING TRIGGER, AND THEN SIMULTANEOUSLY PRESS THE ANVIL RELEASE SWITCH ON EITHER SIDE OF THE INSTRUMENT. WHILE PRESSURE IS STILL ON THE ANVIL RELEASE SWITCH, SLOWLY RELEASE THE CLOSING TRIGGER. THE DEVICE WAS DISASSEMBLED TO RESET THE BAILOUT SYSTEM AND THEN, THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE; NO ABNORMAL NOISE WAS NOTED. EVENT COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DIFFICULTIES NOTED DURING THE FUNCTIONAL TESTING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A THORACIC LOBECTOMY, WHEN FIRING OVER THE BRONCHUS, THE BLADE GOT STUCK. THE KNIFE REVERSE WOULD NOT WORK. THE MANUAL OVER-RIDE WAS USED TO OPEN THE JAWS. IT WAS A PARTIAL FIRE. THE DEVICE WAS NOT FIRED OVER AN EXISTING STAPLE LINE OR CLIP. THIS WAS A RELOAD FIRING, BUT IT IS UNKNOWN WHAT FIRING THEY WERE ON. SAME LIKE DEVICE WAS USED TO COMPLETE THE CASE. NO PATIENT CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86916 PWRD ECH FLEX 60MM STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1